Nurse Call System Compliance Standards: UL 1069, NFPA 99, UL 2560 and AS 3811 Explained
By Nurse Call Technology Editorial Team · 2 October 2026

Understand how UL 1069, NFPA 99, UL 2560 and AS 3811 define nurse call system compliance requirements for hospitals and aged care. Read the full breakdown.
Nurse Call System Compliance Standards: The 2026 Guide to UL 1069, NFPA 99, UL 2560 and AS 3811
By the Nurse Call Technology Editorial Team
OSHA has said that connecting a certified nurse call system to an uncertified pocket pager can void the certification. That one interpretation shows why nurse call system compliance standards apply to the whole installed system, not just to individual parts. This guide sets out which standards apply in the US and Australia, what each one requires, how often systems should be tested, and gives you an audit checklist for handover and inspections.
Contents- Quick answer: which standards apply
- UL 1069: the US equipment standard
- NFPA 99 and NFPA 70: the installation codes
- OSHA and NRTL certification
- UL 2560 for assisted and independent living
- The Australian standard: AS 3811
- Core nurse call system specifications
- How often to test a nurse call system
- Compliance audit checklist
- Common compliance mistakes
- FAQ
Key facts
Edition 8 (2024)UL 1069, the Standard for Hospital Signaling and Nurse Call Equipment, is currently in its 8th edition, published and ANSI-approved on February 8, 2024. Source: UL Standards (ULSE) (February 8, 2024)UL 1069 states that a fundamental nurse call system (NCS) must conform to the minimum requirements established in NFPA 99 (Health Care Facilities Code) and NFPA 70 (National Electrical Code).Source: UL Standards (ULSE) (February 8, 2024)Under UL 1069, requirements for nurse call systems do not cover medical and dental equipment, line isolation monitors, X-ray equipment, and certain other specialized hospital equipment covered by separate standards.Source: UL Standards (ULSE) (February 8, 2024)OSHA has stated that nurse call systems are tested against the provisions of UL Standard 1069, or its equivalent, and this testing forms the basis for certification or approval by a Nationally Recognized Testing Laboratory (NRTL), which OSHA requires for such equipment.Source: U.S. Occupational Safety and Health Administration (OSHA) (December 29, 1989)Per OSHA guidance, if a nurse call system interfaces with a pocket paging system, the whole combined system must be tested against UL 1069 or its equivalent, otherwise the NRTL certification is not valid and will not meet OSHA requirements.Source: U.S. Occupational Safety and Health Administration (OSHA) (December 29, 1989)UL Solutions publishes two distinct standards for these systems: UL 1069 for hospital signaling and nurse call equipment, and UL 2560 for emergency call systems used in assisted living and independent living facilities outside the hospital setting.Source: UL Solutions (2024)Quick Answer: Which Nurse Call System Compliance Standards Apply?
In the United States, nurse call system compliance standards are built on three layers. UL 1069 covers the equipment. NFPA 99 and NFPA 70 cover performance and installation. OSHA requires certification by a Nationally Recognized Testing Laboratory (NRTL). Assisted living uses UL 2560 instead. In Australia, AS 3811 is the reference standard for nurse call design and performance.
Each layer answers a different question:
- Equipment standard (UL 1069, UL 2560, AS 3811): what the hardware and software must do.
- Facility and electrical codes (NFPA 99, NFPA 70): how the system behaves in a clinical environment and how it is wired.
- Regulatory enforcement (OSHA, state health departments, accreditation surveyors): who checks compliance and what evidence they accept.
For background on how nurse call fits into hospital signaling more broadly, see our pillar guide, what is nurse call and how hospital signaling equipment is regulated.

UL 1069: The US Equipment Standard for Nurse Call
UL 1069 is the Standard for Hospital Signaling and Nurse Call Equipment. It is the central US standard among nurse call system compliance standards. The current 8th edition was published and ANSI-approved on February 8, 2024.
According to UL Standards, a fundamental nurse call system must conform to the minimum requirements in NFPA 99 and NFPA 70. In practice, a UL 1069 listing means the product was evaluated against those codes as installed. UL 1069 also has clear limits on what it covers. It does not apply to medical and dental equipment, line isolation monitors, X-ray equipment, or certain other specialized hospital equipment, which have their own standards.
UL Solutions describes UL 1069 as a performance-based standard that ensures minimum system configurations are met and calls are properly routed to the nurse station. Facilities that need to show NFPA 99 compliance usually look for a UL 1069 listing.
NFPA 99 and NFPA 70: The Installation Codes
NFPA 99, the Health Care Facilities Code, sets how nurse call signals must behave in a clinical environment. NFPA 70, the National Electrical Code, governs the wiring and power. Together they set the installation requirements that UL 1069 equipment must meet.
The NFPA 99 emergency call requirement is the one most often checked in the field. Under the 2012 edition, each nurse call station must initiate a distinct visual and audible emergency signal that can only be cancelled at that same station. The signal must activate an annunciator at the nearest nursing station and a visual signal outside the patient's room. This design ensures staff have to physically attend the room to clear an emergency.
Facilities that adopt a later NFPA 99 edition should check the current text, because clause wording and numbering change between editions. The design intent behind each rule is more useful to apply than an old clause number.
OSHA and NRTL Certification
OSHA requires nurse call systems to be certified by a Nationally Recognized Testing Laboratory. That certification is based on testing against UL 1069 or an equivalent standard. Of all nurse call system compliance standards, NRTL certification is the one with the most direct federal enforcement behind it.
The integration rule matters most. OSHA's interpretation states that if a nurse call system interfaces with a pocket paging system, the whole combined system must be tested against UL 1069 or its equivalent. If it is not, the NRTL certification is not valid. That 1989 guidance referred to pagers, but the same reasoning applies to the phones, wearables, and middleware that facilities connect today.
- Ask for a listing that names the exact integration path, not just the base nurse call controller.
- Record software and firmware versions at handover. An unlisted upgrade can take the system outside its tested configuration.
- Treat every new third-party integration as a compliance change, not just an IT change.
UL 2560 for Assisted and Independent Living
UL 2560 covers emergency call systems in assisted living and independent living facilities outside the hospital setting. UL 1069 remains the standard for hospital nurse call. UL Solutions publishes both standards separately.
Mixed-use campuses need particular care. A skilled nursing wing may require UL 1069, while an independent living tower on the same site falls under UL 2560. Choosing one platform for the whole campus only works if it holds the correct listing for each occupancy type. Aged care retrofits often rely on wireless nurse call systems designed for retrofits and aged care. For these, confirm that the wireless components are covered by the listing.
The Australian Standard: AS 3811
AS 3811 is the Australian Standard for nurse call systems. It covers call types, prioritisation, annunciation, and system reliability. It is the reference point for nurse call system compliance standards in Australian hospitals and residential aged care. State health departments often add their own engineering and ICT specifications on top of it.
Australian public health projects usually specify minimum workflows beyond basic calls:
- Nurse call (patient-initiated)
- Nurse presence (staff registering in the room)
- Staff assist
- Emergency
- Code blue / medical emergency
Australian projects should be designed to the edition of AS 3811 named in the contract and to the relevant state health facility guidelines. Aged care providers should also be able to show how call response data supports their quality reporting obligations.
Core Nurse Call System Specifications
A compliant nurse call specification defines the call types, the signals at the room, corridor, and station, the rules for cancelling calls, how calls are escalated, supervision of system faults, and the integration paths. The table below compares how the major nurse call system compliance standards treat each area.
| Area | UL 1069 + NFPA 99 (US hospitals) | UL 2560 (US assisted living) | AS 3811 (Australia) |
|---|---|---|---|
| Scope | Hospital signaling and nurse call | Emergency call outside hospitals | Hospitals and aged care nurse call |
| Emergency cancel | At originating station only (NFPA 99, 2012) | Defined in standard; check edition | Defined in standard; check edition |
| Annunciation | Nurse station plus outside-room visual | Routing to responsible staff | Prioritised display and audible cues |
| Certification proof | NRTL listing (OSHA) | NRTL listing | Manufacturer declaration plus commissioning tests |
| Integration risk | Combined system must be tested | Same principle applies | Specified per project ICT spec |
Infrastructure choice affects how easily you can prove compliance. Hard-wired nurse call systems for hospitals usually include line supervision. Wireless systems need documented battery and signal monitoring instead.
How Often Should a Nurse Call System Be Tested?
Test the system fully at commissioning. After that, follow the manufacturer's maintenance schedule and your facility's risk-based policy, which should include routine functional checks of every call point. Nurse call system compliance standards define what the system must do. Local policy, accreditation bodies, and the manufacturer usually set how often it is tested.
A workable testing structure has four tiers:
- Commissioning: every device, call type, cancellation rule, and integration path, recorded against the specification.
- Routine functional checks: rotating rounds of call points, pull cords, and bathroom emergency points.
- Preventive maintenance: controllers, power supplies, batteries, and fault supervision, on the manufacturer's schedule.
- Change-triggered retests: after any firmware update, new integration, or room reconfiguration.
Worn cords and damaged call points are the most common failures found during rounds. Keeping spare call points, annunciators, and accessories on hand lets you fix them the same day.
Nurse Call Compliance Audit Checklist
This checklist from the Nurse Call Technology Editorial Team turns nurse call system compliance standards into evidence that a surveyor or engineer can verify. Use it at handover and at each annual review.
- Applicable standard and edition for each occupancy (UL 1069, UL 2560, or AS 3811)
- NRTL listing covering the exact installed configuration, including integrations
- Firmware and software version register, matched to the listing
- Commissioning test sheets for every room and call type
- Proof that emergency calls cancel only at the originating station
- Fault supervision test records (power, cable, wireless link, battery)
- Routine functional check logs with corrective actions closed
- Change log linking each modification to a retest
To score a candidate platform against these points before you buy, use our guide to choosing, costing, and proving a nurse call tool works.
Common Nurse Call Compliance Mistakes
Most failures against nurse call system compliance standards come from changes made after commissioning, not from the original design. These are the mistakes we see most often:
- Unlisted integrations: connecting phones or middleware without combined testing. Under OSHA's interpretation, this invalidates the NRTL certification.
- Edition mismatch: specifying to the latest UL 1069 edition when the state code references the 7th, or the reverse.
- Wrong standard for the occupancy: installing UL 2560 equipment in a space that requires UL 1069.
- Remote cancellation: configuring emergency calls so they can be cleared from the station, which conflicts with the NFPA 99 design intent.
- No retest after changes: room fit-outs that move or remove call points without updating test records.
If you need help mapping your facility against these requirements, contact Nurse Call Technology for a compliance review or project quote.
Nurse Call System Compliance Standards FAQ
What is the Australian standard for nurse call systems?
The Australian standard for nurse call systems is AS 3811. It covers call types, prioritisation, annunciation, and reliability. State health departments often add their own ICT specifications, which define minimum workflows such as nurse presence, staff assist, emergency, and code blue.
What are the specifications of a nurse call system?
A nurse call specification defines the call types, room and corridor signals, nurse station annunciation, cancellation rules, escalation, fault supervision, and integration paths. In the US, NFPA 99 (2012) requires a distinct emergency signal that can be cancelled only at the originating station.
How often should a nurse call system be tested?
A nurse call system should be fully tested at commissioning. After that, it should be checked on the manufacturer's maintenance schedule and your facility's risk-based policy, and retested after any firmware, integration, or room change. The standards define performance. Local policy usually sets how often testing happens.
What are the 7 nursing standards in Australia?
The seven nursing standards in Australia are the Nursing and Midwifery Board of Australia's Registered Nurse Standards for Practice. They cover critical thinking, therapeutic and professional relationships, capability, assessment, planning, delivering care, and evaluating outcomes. These are professional practice standards, not equipment standards, and they are separate from AS 3811.
Does adding smartphones or pagers affect UL 1069 compliance?
Yes. OSHA guidance states that when a nurse call system interfaces with a paging system, the combined system must be tested against UL 1069 or its equivalent. Otherwise, the NRTL certification is not valid.
Is UL 1069 or UL 2560 required for assisted living?
UL 2560 covers emergency call systems in assisted and independent living outside hospitals. UL 1069 covers hospital nurse call. Mixed-use campuses may need equipment listed to both, depending on each occupancy.